 NOTE:
After setting up your most common regimes as a protocol set up the
next or last protocol letter as a 'Tailored' or one off regime that you
can modify for individual patients who don't fall within the pre-set
protocols. Once the infusion has been administered for a particular
patient you can go back in and change this for the next or a future
patient.
 NOTE:
Each protocol parameters are unique to that protocol. For example if
you use both ropivacaine and bupivacaine in your service and one
commonly comes in a 100 ml bag whilst the other is in a 250 ml bag
than you can set these bag sizes as standard for each.
 NOTE:
Entry of all parameters is via the numeric keypad. Only the protocol
selection is a scroll option menu using the up and down arrows. The
previous value will be displayed. Enter the new value which will
automatically delete the previous value. If you make an error whilst
entering data, pressing the
Repeat as necessary.
To modify protocol
1.
Press
The screen than will prompt for code. Enter
relevant code and press
2.
From the menu select "Modify Protocol" by
scrolling with the arrow keys and pressing
.
3.
The pump will prompt for code.
Software ver. CVPCA30E_1
. The following will appear.
27