Description
Software ver. CVPCA30E_1
Result
Possible Cause
Infusion stops
Fatal internal error has occurred
Program ends,
Current infusion program has
pump will turn to
been completed.
KVO mode
Pump will not
Administration set loaded
start
incorrectly (the key was not
located in its place correctly).
A set different than the
BodySet®/ MicroSet® was
loaded in the pump.
Setting cannot be
Lock Out mode is turned on.
changed
Setting keys do
Lock Out mode is turned on.
not function
Required Action
Restart the pump. If the alarm does
not stop pack the pump properly, in
its original packaging, and send it for
service.
Press STOP/NO to restart a new
program or turn the pump OFF.
Reload the set into the pump as per
the instructions.
Change the set to a BodySet®/
MicroSet® and load it according to
instructions.
No action required. If changing
program is required, unlock the pump
and refer to lock section for further
instructions.
No action required. If changing
parameters is required, unlock the
keys and refer to lock section for
further instructions.