14.
To confirm concentration of mg/ml Enter the
correct value and press
concentration of mcg/ml press
15.
Enter the correct value and confirm with
the relevant units. Use the numeric keypad to enter the value/ Press
when done.
16.
Set the protocol maximum rate. This will set
the maximum rate allowed for titration.
17.
Set the Bolus using the numeric keys and
confirm with
NOTE: BodyGuard® is capable of delivering the bolus dose at up to 1000
ml/hr. The default setting is 300 ml/hr. The user can adjust bolus delivery
down to a minimum of 100 ml/hr in "Change set up mode" using the level
three access code.
NOTE: The BodyGuard® has a special feature which optimizes bolus
speed during infusions through narrow catheters such as PCA catheters.
The pump will attempt to deliver the bolus at the pre-set bolus delivery rate.
If the pump senses a back pressure above the pre-set occlusion pressure
level during bolus delivery it will automatically reduce the bolus delivery
speed until it senses a pressure below the level at which an occlusion
alarm would normally sound. If it still senses an occlusion pressure above
this level at the minimum bolus delivery rate the pump will then alarm
'Down Occlusion'. This feature ensures optimum bolus delivery speed and
reduces false Occlusion alarms during bolus delivery through narrow
catheters such as PCA or soaker catheters.
Software ver. CVPCA30E_1