Loading the Administration Set

NOTE: Follow the instructions supplied with the individual administration set.
Warning: Use of any administration set, other than the BodySe/MicroSet®,
may impair the operation of the pump and the accuracy of the
Infusion.
Lift the door latch (on the right side of the door) to open the BodyGuard® pump
1.
door and remove the administration set from the sterile packaging leaving the
end caps on the line.
Connect the administration set to the medication bag.
2.
Hold the section of tubing with the black key (small plastic block) and make sure
the flow direction is in line with the flow direction arrows inside the pump door.
3.
Insert the IV tubing into the pump by placing the key into the keyway as shown
by the arrow in the diagram. Insert the tubing from left to right, and avoid
stretching or pulling the tubing. Check that the key located on the tube is placed
in its correct position in the tubing guide.
NOTE: The key can only be fitted into the key space one way. If you are having
troubles fitting it do not force it in. Try to turn the line around to ensure
you have correctly lined up with the direction of flow.
4.
For products equipped with Anti-Stretch locator, locate the round blue Anti-
Stretch locator in the round space on the left side.
5.
Verify that the distal flow valve is on the right hand side of the pump.
6.
Close the pump door until the catch clicks.
NOTE: Ensure that the tubing is inserted completely into the pumping channel.
Warning: Use aseptic technique. Patient infection may result from the use of
non-sterile
components and use only once single use consumables marked
Software ver. CVPCA30E_1
components.
.
Maintain
sterility
24
of
all
disposable