6 – Specifications
Pumping Mechanism:
Flow Rate:
Priming Rate:
Total Infused Volume:
Total Time Setting:
Accuracy:
KVO rate:
Air Sensor:
Maximum Pressure:
Power Supply:
Battery:
Battery Operation
Battery Charging:
Alarms:
Software ver. CVPCA30E_1
Piston Pump
0 to 100 ml/hr in 0.1ml increments
300ml/h to 425ml/h (default 400ml/h)
0.1 to 1,000 ml.
100 hours
±5%
0.1 to 5 ml/hr
Ultrasonic, adjustable air bubble size
10 psi Adjustable (250mmHg – 520mmHg
(5psi – 10psi )).
110-240 VAC, 50/60 Hz. 0.3A max.
Rechargeable Li-Ion 7.4V, 1800mAh. (Typical)
17 hours at 100ml/hr (Rechargeable)
Automatic when clicked into the Charger that is connected to an
AC power source. Six hours needed to charge a fully-depleted
battery.
When a problem is detected, the BodyGuard® displays the
following alarms:
Air in line
Down Occlusion
Pump Unattended
End of infusion
Low Battery
End Battery
Door Open
Error (followed by error code)
Lock Mode
Lock Out Mode
Missing Key
51

6-Specifications