Charging The Pump - CME BodyGuard 575 Color Vision Operator's Manual

Pca infusion system, pain manager
Table of Contents

Advertisement

Charging the Pump

NOTE:
Warning:
1. Connect the charger unit to AC power and verify that the AC indicator is lit.
2. Put the pump into the charger (top first) until it clicks into place.
.
3. The battery is charging when the red battery LED on the front of the charger is lit.
The battery is fully charged when the battery LED symbol turns green. If the led
does not light or does not turn green after more than 6 hours, pls. replace battery.
4. Remove the pump from the charger by pressing the release button (located next
to the LED indicators on the charger), Lift the pump out of the charger bottom first.
Caution: Leaving the battery discharged for a long period of time may damage
the battery.
Software ver. CVPCA30E_1
The pump is protected against overcharging. In order to keep your
battery fully charged, connect the pump to the mains via the
Charger whenever possible. Be sure that the battery is fully
charged at all times.
If the battery is removed during operation while pump is not
connected to a charger connected to mains, the pump will turn
off.
15