Charging The Pump; Operation; Before Operating The Pump - CME BodyGuard 121 Twins Operator's Manual

Infusion system
Table of Contents

Advertisement

Charging the Pump

NOTE:
The pump is protected against overcharging. Connect the pump to the mains via the
charger whenever possible to be sure that the battery is fully charged at all times.
Warning:
disconnected from mains, the pump will turn off.
1. Connect the charger unit to AC power, and verify that the AC indicator is lit (charger right side above
channel II).
2. Put the pump into the charger and tighten the safety screw. Please see drawing on the charger.
3. The battery is charging when the red battery LED on the front of the charger is lit. (charger lower left
side above channel I) The battery is fully charged when the full battery LED symbol turns red
(charger upper left side above channel I).
4. Remove the pump from the charger by releasing the safety screw and lifting the pump out of the
charger.
Caution:
4 – Operation

Before Operating the Pump

Before attaching the system to a patient, run the following tests to verify that all indicators and
alarms work properly. Perform the test on both channels.
When an alarm is activated the following occurs:
An alarm message appears on the pump and charger display. Each channel has its own display.
The alarm message will be shown on the display relevant to the channel in which the alarm
occurred.
An audible alarm sounds.
Infusion ceases.
The operation LED changes from green to red.
NOTE:
All tests should be performed in the Continuous program on both channels.
If the battery is damaged during operation, while pump and charger are
Leaving the battery discharged for a long period of time may damage the battery.
20