6-18. Excitation Voltage

This function selects the bridge excitation voltage to be supplied to the strain
gauge sensor.
65
Setting Excitation Voltage
Select setting mode 3.
1)
+ -
FNC
ZERO
Select excitation voltage.
2)
Press eight times.
CAL
HOLD
Use
and
GAIN
then use
key to validate the setting.
HOLD
To return to the indicated value display, press
Use a strain gauge sensor to be connected to the F320 whose maximum
excitation voltage is above the bridge excitation voltage specified.
If the bridge excitation voltage is greater than the maximum excitation
voltage of the sensor, the sensor may overheat or may be damaged.
ZERO
Excitation Voltage
1 : 10 V
0 : 2.5V
keys to set the excitation voltage,
CAL
CAUTION
key.
+ -