11-2. Current Output (DAI Option)

This converter is used to obtain an analog output which is linked with the
indicated value of the F320.
The range of the analog output is from 4 to 20mA.
An analog output from 4mA to 20mA can be obtained with respect to any digital
values set by the D/A zero setting and the D/A full scale setting functions.
The output circuit and the main circuit are isolated.
Current output:
D/A conversion speed:
Resolution:
Over range:
Zero adjustment range: Approx. 3.6 to 4.4mA (at the time of zero output)
Gain adjustment range: Approx. 19.6 to 20.4mA
Zero drift:
Gain drift:
Non-linearity:
Output connector:
Current output terminals
Terminals for obtaining current signal.
"+" is for signal, and "-" is for ground.
4 to 20mA can be obtained.
+
-
Zero adjustment trimmer
Trimmer to adjust so that the current output becomes
zero when the indication is as registered by the D/A
zero setting function.
The zero adjustment range is approx. 3.6 to 4.4mA
(at the time of zero output).
4 to 20mA (load resistance; 350Ω or less)
2000 times/sec.
1/10000
3.2 to 20.8mA
(at the time of full scale output)
Within 0.5µA/ ℃
Within 50ppm/ ℃
Within 0.05%FS
* Not including the drift of the analog input section
Screw type terminal block (two-pole)
11.D/A CONVERTER
Gain adjustment
trimmer
Trimmer to adjust so
that
the
indication
becomes as registered
by the D/A full scale
setting function.
The gain adjustment
range is approx. 19.6
to 20.4mA (at the
time of full scale
output).
94