Preparing The Device; Connecting To The Mains Power Supply - Atmos S 351 NATAL Operating Instructions Manual

Hide thumbs Also See for S 351 NATAL:
Table of Contents

Advertisement

Device suspension for Serres®

3.2 Preparing the device

Read the safety information in chapter „2 Notes for your safety" on page 12 carefully
before using the product.
Damaged pump diaphragms due to cold temperatures during transport.
1. In case the device was transported at temperatures below −5 °C: let the device stand
at room temperature for at least 6 hours before proceeding with the next steps
2. Check the device, secretion canisters, power cable, accessories and hoses for possi-
ble damage.
3. If the device is damaged: document and report the damages in transit. Send the
device to ATMOS (chapter „6.3 Sending in the device" on page 41).
4. If the device is not damaged: place the device on a safe, level surface.

3.3 Connecting to the mains power supply

1. Check whether the voltage and frequency data listed on the device correspond to
the values of the mains power supply.
2. Connect the device to the mains.
3. Secure the power cable against accidental removal by using the safety clamp.
18 Setting up and starting up