Control panel
5
Display of the battery status
The following display values are not valid during battery charging.
≤ 100 % ca. 70 %
The battery status must be supervised visually.
If only the red LED lights up, although the battery has been loaded for a longer period,
the battery is defective.

3.2 Preparing the device

Prior to first operation peruse the safety notes in chapter „2.0 Hints for your safety" on
page 11.
Damaged pump diaphragms due to cold temperatures during transport.
1. After transport in temperatures below -5°C: The device must be acclimatized for up
to 6 hours at room temperature before you continue with the next steps.
2. Check the device for any damage in transport.
3. If the device is damaged: Document and report the transport damage. Send in the
device to ATMOS (chapter „6.3 Sending in the device" on page 40).
4. If the device is not damaged: Place it on a safe and even surface.
5. Check the charging accessories for any damage.
6. Replace defective charging accessories immediately.
7. The battery must be fully charged, chapter „3.3 Charging the battery" on page 17.
8. Remove the canister system from the support.
9. With DDS canister system: Prior to first use, clean the canister system and insert
a bacterial filter (chapter„5.0 Cleaning and disinfection" on page 32 as well as
chapter „3.4 Connection and removal of canister system and hoses" on page 18).
10. Connect the suction hose.
16
Setting up and starting up
4
3
ca. 40 % < 10 %
Display of battery status
1
Vacuum controller
2
Vacuum display
3
1
2
On/off button
4
Power indicator
5