Atmos S 351 NATAL Operating Instructions Manual page 48

Hide thumbs Also See for S 351 NATAL:
Table of Contents

Advertisement

Protective earth conductor
resistance
Earth leakage current
Housing leakage current
Patient leakage current
Heat output
Noise level
Ambient conditions for
transport/storage
• Temperature
• Humidity without conden-
• Pressure
Ambient conditions for
operation
• Temperature
• Humidity without conden-
• Pressure
Max. operational altitude
Contamination level
Overvoltage category
Dimensions H x W x D
Weight
Periodic tests
Protection class (EN 60601-1)
Degree of protection
Type of protection
Further classifications accord-
ing to other regulations
Risk class (according to MDD)
Risk class (according to MDR)
CE marking
GMDN code
UMDNS code
MD/MDS code
ID no. (REF) basic devices
Basic UDI device identifier
48 Technical data
sation
sation
max. 0.1
max. 0.5 mA
max. 0.1 mA
max. 0.1 mA
approx. 135 J/s
< 54 dB (A) @ 1 m (according to ISO 7779)
−10...+60 °C
30...95% air humidity without condensation
at an air pressure of 700...1060 hPa
+5...+40°C
30...95% air humidity without condensation
at an air pressure of 700...1060 hPa
3000 m (above MSL)
Class 2
III
Without trolley: 30 x 33 x 20 cm
With trolley: 84 x 49 x 52 cm
10.2 kg (without secretion canister and without trolley)
Repeat test of electrical safety every 12 months.
Recommended: inspection according to manufactur-
er's specifications.
I
Applied parts type B
IPX0
Class IIa according to Rule 11
Class IIa according to Rule 12
63643 Surgical suction pump
14-317 Vacuum extractor, obstetrical
MDA 1104 Active surgical device
MDA 0318 Other active non-implantable devices
444.0401.0 (230 V)
444.0401.1 (100 V)
444.0401.2 (115 V)
444.0401.3 (127 V)
42503651VEUnitS351NATALTN