Pre-cleaning
Flushing:
5x / 30 s
Disassembly
Mechanical cleaning
and disinfection
Pre-rinse:
1 min
Clean:
5 min
55 °C / 131 °F
Neutralise:
2 min
Disinfect:
5 min
93 °C / 199 °F
Dry:
12 min
110 °C / 230 °F
Checking and
maintaining
Assembly
Function check
Packaging
Sterilisation
Prefractionated
3x
vacuum:
Temperature:
134 °C / 273 °F
Time:
5 min
Dry:
10 min
Storage
 Pre-cleaning is only necessary for mechanical cleaning
and disinfection.
1. Clean the accessories evenly under running water.
2. Rinse the hollow spaces and lumina of the accessories
thoroughly under running water.
Not necessary.
1. Empty the secretion canister.
2. Clean and disinfect using a suitable programme:
• Pre-rinse with cold water
• Cleaning with cleaning agent
• Neutralise with cold water
• Intermediate rinse with softened, cold water
• Disinfection with suitable disinfectant and demineral-
ised water
• Drying
Cleaning and disinfection
device:
Programme:
Adapter:
1. Check whether reprocessing was successful with a
suitable light magnifier.
2. If reprocessing was unsuccessful, reprocess the accesso-
ries again.
3. Dispose of damaged accessories or have them
repaired.
Not necessary.
Not necessary.
1. Label the accessories.
2. Pack the accessories using a packaging system according
to DIN EN ISO 11607.
1. Sterilise the accessories using a suitable procedure:
• Steam sterilisation / autoclaving
 Ideally, always use the same procedure.
Steriliser:
1. Observe the ambient conditions; see chapter „11 Techni-
cal data" on page 47.
• In accordance with EN ISO
15883-1
• Miele Vario TD
• Miele E336/E446
• In accordance with EN 285