Secretion Canister System - Atmos S 351 NATAL Operating Instructions Manual

Hide thumbs Also See for S 351 NATAL:
Table of Contents

Advertisement

Cleaning agent
(manufacturer)
Active ingredients in 100 g
Disinfectants – Manual reprocessing
Gigasept® FF neu (Schülke
<5% phosphonates, <5% anionic surfactants, <5%
& Mayr)
non-ionic surfactants, perfumes, methylisothiazoli-
nones
Cleaning agent
(manufacturer)
Active ingredients in 100 g
Cleaning agents – Mechanical reprocessing
neodisher® MediClean forte
<5% non-ionic and anionic surfactants, enzymes
(Dr. Weigert)

5.4.3 Secretion canister system

Characteristics
The accessories have the following hard-to-reach areas:
• Double hose connector (lumina)
• Secretion canister lid complete (hollow spaces)
Take particular care when reprocessing hard-to-reach areas.
Pre-treating at the site
of use
Flushing:
60 s
Rinsing:
60 s
Collecting and
transporting
Disassembly
Pre-cleaning
Flushing:
1x / 30 s
Rinsing:
60 s
Brush:
Round brush
• Diameter:
7 / 11 / 15 mm
• Material:
Nylon
• Characteristics: With angled
head
1. Empty the canister.
2. Clean the accessories under cold, running water.
3. Rinse the hollow spaces and lumina of the accessories
thoroughly under running water.
»
No more residue is visible.
1. Label any damaged accessories.
2. Place the accessories in a canister.
3. Transport the canister to the reprocessing site.
See chapter „30" on page 23.
1. Discard disposable products.
 Pre-cleaning is only necessary for mechanical cleaning
and disinfection.
1. Make the following hollow spaces accessible:
• Double hose connector
• Entire secretion canister lid
2. Make the following lumina accessible:
• Double hose connector
3. Clean the accessories evenly with a suitable brush under
running water.
4. Rinse the hollow spaces and lumina of the accessories
thoroughly under running water.
Type
Liquid
concentrate
X
Type
Liquid
concentrate
X
Reprocessing
35