Spanner/wrench
An equipment error has
shows on the
occurred.
display.
Thermometer
Ventilation slits are
shows in the display
blocked.
(ATMOS S 351
NATAL has over-
Ambient temperature is
heated).
too high.
Fan is defective.
Only a low vacuum
A low target vacuum level
level can be set with
has been set with the
the foot controller.
buttons.
The device does not
The foot controller was
recognise the foot
connected after switch-
controller.
ing on the device.
If the symbol disappears, the device
was able to correct the error. Nev-
ertheless, have your suction device
checked by ATMOS Service.
Only emergency operation is possi-
ble as long as the symbol shows. Call
ATMOS Service.
Check the ventilation slits (bottom
of the device); they must not be
blocked.
Use the ATMOS S 351 NATAL only in
the specified temperature range. Try
using auto standby (less heat build-
up).
Call ATMOS Service.
Set the target vacuum to a higher
value (or MAX) using the + button to
obtain a larger control range for the
foot controller.
Connect the foot controller to the
device before starting it.
Troubleshooting