4.11 Changing the secretion canister

Change or empty the secretion canister when it is 2/3 full.
WARNING
Risk of infection.
Death or serious injuries from infection.
 Always wear disposable gloves when changing the secretion canister.
1. Stop the suction process by switching off the pump.
2. Remove the double hose connector from the full secretion canister.
3. If a second secretion canister is attached, insert the double hose connector into it.
4. Lift the full secretion canister with its lid upwards. Press the release button and open
the locking handle.
5. Empty the secretion canister or replace it with a new one.
 Dispose of the suction material properly.
6. Attach the canister lid to the empty secretion canister and then to the device.
7. Insert the double hose connector into the secretion canister.

4.12 Warning messages

Display
28 Operation
Cause
Bacterial filter is blocked.
 This warning message
also appears if there is
a kink in the hose or if
drainage accessories
are connected.
Short circuit between the
contact terminals.
The secretion canister is
full.
Troubleshooting
Replace the bacterial filter.
Check the contacts on the
device.
Change the secretion
canister.
In the event of strong
foaming:
Fit the anti-foaming device
over the sensor.
Please observe:
The warning message
'Secretion canister full' is
deactivated when dispos-
able canister systems are
in use.