12 Notes on EMC

 Medical electrical devices are subject to special precautions with regard to EMC and
must be installed according to the following EMC notes.
Guidance and manufacturer's declaration – ambient conditions
The product is suitable for use in the following environments:
• Home care in any building, outdoor areas and means of transport.
• In professional healthcare facilities such as medical practices, clinics, first-aid facili-
ties and operating theatres/rooms.
It is not suitable for use in the vicinity of HF surgical devices and in settings outside
of an HF-shielded room of a magnetic resonance imaging system.
• Special environments such as factory or military facilities and medical areas near HF
surgical devices, short-wave therapy equipment or within an HF-shielded room of a
magnetic resonance imaging system.
The customer or user must ensure that the device is used in a prescribed environment.
Guidance and manufacturer's declaration – key features
 Please note the technical data in this manual. The essential features are fully usable
even in the presence of electromagnetic disturbances.
Guidance and manufacturer's declaration – electrical components
The product has the following electrical components:
Type
Power cable with low-heat device plug
IEC 60320C14
Guidance and manufacturer's declaration – warnings
Warning
The use of electrical components and accessories other than those specified or provid-
ed by the manufacturer may cause increased electromagnetic interference or reduced
immunity to electromagnetic interference and result in faulty operation of the device.
Warning
Portable RF communications equipment (e.g. radios, antenna cables) should not be
used within 30 cm* of any parts or cables of the product as specified by the manufactur-
er. Failure to heed this warning may result in a reduction in the device's performance.
 *The distance may be reduced at higher immunity test levels.
Warning
Avoid placing the device on top of or next to another device. This could result in faulty
operation. If such placement cannot be avoided, the proper functioning of the device
must be monitored regularly. If possible, please switch off any nearby devices that are
not in use.
REF
Max. cable length
008.0629.0
5 m
Notes on EMC