6 Maintenance and service

Maintenance, repairs and periodic tests may only be carried out by persons who have
the appropriate technical knowledge and who are familiar with the product. The per-
son must possess the test devices and original spare parts required to carry out these
measures.
ATMOS recommends: work should be carried out by an authorised ATMOS service
partner. This ensures that repairs and tests are carried out professionally, original spare
parts are used and warranty claims remain unaffected.
Maintenance, repairs and periodic tests may not be carried out while the product is
being used on the patient.

6.1 Periodic tests

Perform a repeat test of the electrical safety according to IEC 62353 at least every 12
months.
In this context, ATMOS recommends conducting an inspection in accordance with the
manufacturer's specifications.

6.2 Function check

6.2.1 Visual inspection

• The power cable shows no signs of damage
• The housing is undamaged
• If available: the device is connected correctly and securely to the trolley
• If available: the foot controller is connected correctly
• A new bacterial filter is installed
• The accessories are complete and without damage