Dräger Atlan A300 Instructions For Use Manual page 276

Hide thumbs Also See for Atlan A300:
Table of Contents

Advertisement

Technical data
16.13
Measuring systems and displays (continued)
Resolution of displayed value
Pressure gauge for indicating the pressure
in the internal breathing system
Range
Accuracy
Resolution of displayed value
Volume
Tidal volume
Expiratory
Range
Accuracy for gas measurement with PGM
Accuracy without PGM
Resolution of displayed value
Difference between inspiratory and expira-
tory tidal volumes
Range
Accuracy for gas measurement with PGM
Accuracy without PGM
Resolution of displayed value
Minute volume
Total
Mandatory
Spontaneous
Range
Accuracy for gas measurement with PGM
Accuracy without PGM
Resolution of displayed value
276
1 hPa (cmH
–20 to +80 hPa (cmH
±5 % of the measured value or ±2 hPa
(cmH
5 hPa (cmH
The measured volume values displayed
already take account of the compliance of the
breathing circuit. As soon as CO
phases are detected, the sample gas flow for
the patient-gas measurement module is addi-
tionally taken into account.
VT
0 to 2500 mL
±10 % of the measured value or ±15 mL (the
larger value applies)
±20 % of the measured value or ±40 mL (the
larger value applies)
1 mL
∆VT
0 to 2500 mL
±20 % of the measured value or ±30 mL (the
larger value applies)
±40 % of the measured value or ±80 mL (the
larger value applies)
1 mL
MV
MVmand
MVspon
0 to 40 L/min
±10 % of the measured value or ±0.1 L/min
(the larger value applies)
±20 % of the measured value or ±0.2 L/min
(the larger value applies)
0.01 L/min (MV <1 L/min) or 0.1 L/min
(MV ≥1 L/min)
|
Instructions for use
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
O)
2
O)
2
O) (the larger value applies)
2
O)
2
respiratory
2