Reprocessing
Unscrew the sample line (1).
2. Remove the breathing bag (2) together with the breathing bag hose.
3. Remove the breathing circuit (3) and the filters at positions (4) or (5).
13.6.2
Disposing or reprocessing of patient-specific accessories
1.
Risk of injury to patients due to failure of accessories
Disposable products have been designed, tested, and manufactured exclusively
for single use. Reuse, reprocessing or sterilization can result in failure of the
accessory and injury to the patient.
► Do not reuse, reprocess, or sterilize disposable products.
Dispose of the disposable accessories:
– Sample line
– Breathing circuit
– Breathing bag
– Filter
2. Reprocess the reusable accessories in accordance with their instructions for
use.
– Breathing circuit
– Breathing bag
228
CAUTION
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n