Getting started

6
Getting started
6.1

Safety instructions

Checking the safety systems
Some safety systems are only checked during start-up. If this check is not
performed regularly, a device malfunction may occur, putting the patient at risk.
► Restart the device at least once per month so that the safety systems will be
checked regularly.
6.2

Turning on the device

Prerequisite: The device has been reprocessed (see page 222) and assembled
ready for operation (see page 45).
1. Connect the device to the mains power supply.
2. Set the main switch to position I.
3.
Risk of device malfunction
Condensed water may form when the device is brought from a cold storage
location into a warm environment.
► To prevent condensation and resulting failures of electrical components, do
Press the
The device starts. The Standby page is displayed.
If there is sufficient battery charge, the device will also start without the power plug
being plugged in.
6.3

Checking the device configuration

The device can be customized to suit the requirements of the user. Settings for the
following features are possible:
– Start settings for the ventilation
– Alarm limits
– General device behavior
86
1
CAUTION
not turn on the device after abrupt temperature changes for 1 to 2 hours.
key (1).
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n