7.9.4
Deactivating the automatic xMAC monitoring
If the xMAC low alarm (4 or 7) is acknowledged with the ALARM RESET button (5
or 8), the monitoring is deactivated. This prevents renewed alarms as a result of the
anesthetic gas concentration continuing to fall at the end of anesthesia (8). If the
anesthesia is continued (5), the monitoring will be automatically reactivated as soon
as the inspiratory xMAC value rises above the expiratory xMAC value (6).
In CBM mode, the lower alarm limit is adjusted so that no alarm is issued during this
time. Similarly, the value is not limited to 1.0 during this time.
7.10

Changing the patient data

Patient data can be changed during operation.
1. Open the Patient dialog.
2. Modify the patient data.
Changes influence the therapy suggestions, among others, which is indicated by
the position of the arrow
The current therapy settings remain unaffected. Observe the following information:
"Influence of patient category, weight, and age on device behavior", page 311.
When the patient category is changed, the age, weight, and height are
automatically adapted so that they remain within the described limits, see
page 172.
7.11

Exporting data

This device enables the export of screen contents, trends, and data on a USB mass
storage device.
7.11.1
General information
Prerequisite: USB mass storage device is connected to the USB port.
During a saving process, the button turns dark green.
The data are stored in the "Draeger\ExportData" directory.
|
Instructions for use
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Patient
on the therapy controls.
Operation
119