Operation
Device malfunction during operation
Device malfunctions can compromise the correct therapy functionality of the device.
As a result, the patient could be put at risk.
► Only operate the device under the constant supervision of users.
► Always keep a manual resuscitator ready.
Displays and therapy decisions
The status display serves as the primary information source for the internal airway
pressure and for the flow (only with a mechanically controlled gas mixer).
► In the following cases, use the pressure gauge display and the total flow tube
(only with a mechanically controlled gas mixer) as the information source for
therapy:
– The status display has failed.
– The values on the status display do not match the values on the total flow
The following situations could result in misdiagnoses, which could put the patient at
risk:
– Measured values are misinterpreted.
– Measured values are incorrectly or inaccurately displayed.
– The accuracy of the measurements of flow and volume may be impaired if the
breathing system warmer is switched off.
– The parameter under consideration is not meaningful (e.g., MV×CO
uptake, anesthetic agent uptake)
► Do not make therapeutic decisions based solely on individual measured values
and parameters.
► Only use the trend curve for the MV×CO
for therapeutic decisions.
► Therapeutic decisions must be made solely by the user. For further information
see: "User group requirements", page 11.
► Do not use the virtual flow tubes of the electronically controlled gas mixer alone
when making therapeutic decisions.
90
tube.
Instructions for use
and O
uptake parameters as a basis
2
2
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
, O
2
2