Troubleshooting
12
Troubleshooting
12.1
Leakage
Leakage may result in the system not being operational or being ready for operation
with limitations only.
12.1.1
Possible causes of leakage
– The CO
system.
– The APL valve is not correctly fitted to the breathing system or is not set to
30 hPa (cmH
– The breathing bag, the breathing hoses, the Y-piece, or the bacteria filter is
incorrectly fitted or damaged.
– The breathing bag arm (option) is incorrectly fitted to the breathing system. The
sealing ring is soiled or damaged.
– The O-ring on the inspiratory port or expiratory port is damaged, soiled, or
missing.
– The flow sensors are incorrectly installed or damaged. The rear O-ring is
missing.
– The upper part of the breathing system is incorrectly fitted or damaged.
– The breathing system is not locked.
– The valves or seals of the breathing system are damaged.
– The sensor port for the O
– The circuit plug is scratched or damaged.
– The filling or emptying connections on the vaporizer are leaking or are open.
The vaporizer is incorrectly fitted. The O-ring is missing or damaged. The control
dial is not set to the 0 position.
On devices with inspiratory O
– The O
– The O
On devices with integrated patient-gas measurement module:
– The water trap is not connected.
– The sample line is not connected, is kinked, or is leaking.
– The connections for the sample line are damaged.
190
absorber or the CLIC adapter is not securely screwed to the breathing
2
O).
2
sensor is not sealed with the sealing cap.
2
measurement:
2
sensor is not correctly fitted in the breathing system.
2
sensor cell has not been correctly inserted in the sensor.
2
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n