Troubleshooting
Priority
Alarm
!! 255
Inspiratory xMAC high
!! 100
Internal device tempera-
ture high
!
255
Internal device tempera-
ture high
!!! 150
Internal FG flow failure
!
150
Internal FG flow failure
!
60
License expired
214
Cause
The inspiratory anesthetic
gas concentration has
exceeded 3 xMAC for more
than 180 seconds.
A ventilation slot at the rear
of the device is blocked.
Ambient temperature is too
high.
A fan is faulty.
Excessive ventilation set-
tings are applied (e.g., high
respiratory rate, high inspira-
tory pressure, short slopes).
A ventilation slot at the rear
of the device is blocked.
Ambient temperature is too
high.
Excessive ventilation set-
tings are applied (e.g., high
respiratory rate, high inspira-
tory pressure, short slopes).
A fan is faulty.
The internal gas delivery sys-
tem is not operational. A sys-
tem test may be able to
resolve the issue.
The internal gas delivery sys-
tem is not operational. A sys-
tem test may be able to
resolve the issue.
A license has expired. After
next startup, some functions
will no longer be available.
|
Instructions for use
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Remedy
Check vaporizer and fresh-
gas settings.
Check the ventilation slots.
Ensure air flow at the rear of
the device.
Check ambient conditions.
Call Dräger.
Check ventilation settings.
Check the ventilation slots.
Ensure air flow at the rear of
the device.
Check ambient conditions.
Check ventilation settings.
Call Dräger.
Deliver O
: 1. Set the O
2
switch to "Add. O
". 2. Set
2
the O
flowmeter to the
2
desired flow. Check the
vaporizer setting. Make sure
that fresh gas is reaching the
patient.
When the case has been
completed, perform a sys-
tem test.
If the problem persists, call
Dräger.
Set the O
flowmeter to the
2
desired flow. Check the
vaporizer setting. Make sure
that fresh gas is reaching the
patient.
When the case has been
completed, perform a sys-
tem test.
If the problem persists, call
Dräger.
To order a permanent
license, call Dräger.
2