Configuration
Changing the alarm limits
Standby Set-up
I
1 Select Alarm Limits (G) and confirm.
2 Select the alarm limits (H) to be changed and
confirm.
3 Select new value and confirm.
4 If necessary, repeat steps 2 and 3 for other
alarm limits.
5 Select the input arrow (I) and confirm.
The window is closed.
In the following table, the values for the setting
ranges and factory settings are listed for all alarm
limits of Fabius.
Alarm parame-
Setting range Factory setting
ter
O
19 to 100
2
[%]
18 to 99
MV
0.1 to 20.0
[L/min]
0.0 to 19.9
Pressure
10 to 70
[[cmH
O ]
5 to 30
2
(hPa)]
138
H
G
100
20
12.0
3.0
40
8
Changing the minimum alarm volume
Standby Set-up
J
L
1 Select Minimum Alarm Volume (K) and
confirm.
The current minimum alarm volume (L) is
displayed on the screen.
2 Set the new minimum alarm volume to a value
between 1 (minimum) and 10 (maximum) and
confirm.
The window is closed, the cursor is on the input
arrow (J).
Instructions for use Fabius GS premium SW 3.n