Alarm priorities

Fabius assigns the appropriate priority to each
alarm.
The background color of the alarm message field
indicates the priority of the active alarm.
The alarm messages are only displayed on
colored background if the option "Color display" is
activated.
Color
Priority of the alarm message
Red
Warning
Yellow
Caution
Note
The alarm messages are sorted according to these
priorities and displayed corresponding to the
internal priority system. Priority 31 has the highest
and priority 1 the lowest priority. The priority
numbers are listed in the table in chapter ''Alarm –
Cause – Remedy'' on page 157.
There can be a maximum of 4 alarm messages
displayed at the same time in a list. Alarm
messages with higher priority are displayed before
alarm messages with lower priority. Alarm
messages with low priority are only displayed if the
cause for a high-priority alarm is remedied.
*
optional 3 alarm messages for Japan and China
Instructions for use Fabius GS premium SW 3.n
High-priority alarm
Medium-priority alarm
Low-priority alarm
*
Required action
!!!
Immediate action is necessary in order to
avert imminent danger.
!!
Fast action is necessary in order to avert a
danger.
!
Attention is necessary, but a delayed
response is sufficient.
Alarms