Dräger Fabius GS premium Instructions For Use Manual page 159

Hide thumbs Also See for Fabius GS premium:
Table of Contents

Advertisement

Alarm
Alarm
priority
(23/31)
APNOEA FLOW !!!
Instructions for use Fabius GS premium SW 3.n
Cause
The apnea flow alarm based
on a time staggering.
In the Volume Control,
Pressure Control, SIMV/PS
modes with Freq ≥ 6 or in
Pressure Support mode
with apnea ventilation deacti-
vated:
Warning = VT <20 mL for >30
seconds
In the ManSpont, SIMV/PS
modes with Freq < 6 or in
Pressure Support mode
with apnea ventilation acti-
vated:
Warning = VT <20 mL for >60
seconds
Breathing/ventilation
stopped.
Leakage or disconnection in
the breathing system.
Troubleshooting
Remedy
Check ventilator.
Check breathing system.