Troubleshooting
Alarm
Alarm
priority
(23/31)
APNOEA PRESSURE !!
160
Cause
The apnea pressure alarm is
based on a time staggering.
In the Volume Control,
Pressure Control, SIMV/PS
modes with Freq ≥ 6 or in
Pressure Support mode
with apnea ventilation deacti-
vated:
Caution = PAW did not
exceed the pressure thresh-
old value for a duration of >15
seconds.
In the ManSpont, SIMV/PS
modes with Freq < 6 or in
Pressure Support mode
with apnea ventilation acti-
vated:
Caution = PAW did not
exceed the pressure thresh-
old value for a duration of >30
seconds.
Breathing/ventilation
stopped.
Leakage or disconnection in
the breathing system.
Instructions for use Fabius GS premium SW 3.n
Remedy
Check ventilator.
Check breathing system.