Troubleshooting
When the battery is completely empty, Fabius
switches off and generates an acoustic alarm
signal (continuous tone for approx. 30 seconds).
All customized settings, including the alarm limits,
that deviate from the default settings are lost.
The following ventilation modes are still possible:
– Manual ventilation
– Spontaneous breathing
All pneumatic functions of Fabius are still
available:
– APL valve
– Pressure gauge for the airway pressure
– Pressure gauge for gas cylinders and central
gas supply
– Fresh-gas supply and anesthetic agent delivery
– S-ORC
– Flow control valves for O
WARNING
Incorrect patient settings
When the power supply is restored and
Fabius is restarted, all ventilation and alarm
settings are reset to default settings.
After the restart of Fabius, check all settings
and adjust to the patient if necessary.
154
, Air, and N
O
2
2
Instructions for use Fabius GS premium SW 3.n