General information

Units of measurement for pressure
All specified tolerances apply for 20 °C (68 °F),
60 % relative humidity, and 1013 hPa (760 mmHg).
The accuracies indicated below change according
to atmospheric pressure, temperature, and relative
humidity. If one of the ambient conditions is
changed up to the permissible limit, the accuracy of
the corresponding value can change by up to 50
%.If more than one of the ambient conditions are
changed, the accuracy may change by up to
100 %.Example: Accuracy of a pressure measure-
ment value: ±4 % at standard conditions. At 10 °C,
the accuracy changes to ±6 %; at 10 °C and 20 %
relative humidity, to ±8 %.
All patient-related volumes and flow values are nor-
malized to the conditions in the lungs. (BTPS)

Ambient conditions

During operation
Temperature
Air pressure
Relative humidity
Height
During storage and transport
Temperature
Air pressure
Relative humidity
The conditions for use when using additional devices can limit the environment of use of a system as a
whole. Vaporizers and anesthetic agents can limit the use of an anesthesia workstation with regard to its
temperature range and maximum fresh-gas flow. Therefore when using additional devices, follow the
associated instructions for use.
Instructions for use Fabius GS premium SW 3.n
1 hPa = 1 mbar = 1 cmH
100 kPa = 0.1 MPa = 1 bar = 1 kPa x 100
10 to 35 °C (50 to 95 °F)
700 to 1060 cmH
O (hPa)
2
20 to 80 % (without condensation)
Up to 3000 m (9843 ft)
-10 to 60 °C (14 to 140 °F)
700 to 1060 cmH
O (hPa)
2
10 to 90 % (without condensation)
Technical data
O
2
187