NOTE
Fabius contains no components made of natural
rubber latex.
For latex-free use, use a latex-free breathing bag
and breathing hoses.
1 Select suitable accessories for the respective patient category.
Tidal volume
Breathing bag
Breathing circuit
Filter
NOTE
When applying tidal volumes in the range of the
maximum or minimum values indicated for each
patient category, use the smaller breathing bag
and the smaller breathing circuit.
Instructions for use Fabius GS premium SW 3.n
Adults
>700 mL
201 to 700 mL
3 L
Adults
Filter, HMEF, or HME
Pediatric patients
50 to 200 mL
2 L
1 L
Pediatric
Assembly and preparation
Neonates
<50 mL
0.5 L
Neonates
(or pediatric)
Use filters with low
resistance and
compliance.