Connecting the breathing bag

WARNING
Risk of use of toxic or incompatible materials
The breathing bag used must comply with the
current standards.
WARNING
Risk of too high airway pressure or
insufficient fresh-gas
If the breathing bag is pinched, excessive
airway pressures or a lack of fresh gas can
occur.
Attach and position the breathing bag so that
it is not pinched and can inflate freely.
CAUTION
Risk of patient recovering consciousness
A blocked or incorrectly positioned breathing bag
can lead to lack of fresh gas for patients. Manual
ventilation is also not possible.
Make sure that the breathing bag is connected
tightly to the bag holder during the ventilation.
The breathing bag can be mounted to the compact
breathing system with the following variants:
– On a flexible arm
– On a rigid arm
– On the bag elbow directly on the compact
breathing system
Instructions for use Fabius GS premium SW 3.n
Mounting the flexible arm
Prerequisite: Before mounting the flexible
breathing bag holder, the bag elbow on the
compact breathing system must be removed.
D
C
1 Position and align the connection adapter (A)
of the flexible arm on the connection of the
compact breathing system.
2 Tighten the knurled screws (B). Check that the
arm is fixed securely.
3 Fasten the elbow (D) at the end of the flexible
arm.
4 Fasten the breathing bag (C) on the other end
of the elbow.