In the following table, all parameters (B) in
SIMV/PS mode, together with the corresponding
setting ranges and factory settings, are listed.
Parameter
Setting range
Pressure limita-
15 to 70,
tion PMAX
min. PEEP +10
[cmH
O] ([hPa])
and
2
>∆PPS +PEEP
Tidal volume VT
20 to 1100
[mL]
Respiratory rate
4 to 60
Freq [bpm]
([1/min])
Pressure sup-
3 to 20
port ∆PPS
OFF
[cmH
O] ([hPa])
2
PEEP [cmH
O]
0 to 20
2
([hPa])
Trigger sensitiv-
2 to 15
ity Trigger
[L/min]
Inspiratory flow-
10 to 85
Insp Flow
[L/min]
Inspiratory time
0.3 to 4.0
TINSP
[seconds]
Inspiratory
0 to 50
pause
time:inspiratory
time TIP:TI [%]
Instructions for use Fabius GS premium SW 3.n
PAW
Factory set-
ting
40
600
VT
12
10
The respiratory rate Freq defines the time between
0
the individual volume-controlled breaths. The
synchronization of the mechanical breaths is done
with a trigger sensitivity (Trigger) that is activated
2
a specific time before administering a new
mechanical breath: 5 s for respiratory rates (Freq)
below 12 bpm (1/min). For higher respiratory rates,
30
the synchronization is done immediately after the
preceding expiration. Between these mandatory
breaths, the patient can breathe spontaneously.
Mandatory breaths are synchronized with the
1.7
spontaneous breaths of the patient. These
spontaneous breaths can be supported with
∆PPS.
10
The lower alarm limit of the airway pressure PAW
low is used for the airway pressure monitoring to
detect apnea (disconnection) and continuous
pressure.
If the pressure waveform does not cross the
pressure threshold either from above or below, an
alarm is sounded.
PEAK
Trigger
PLAT
∆PPS
Insp Flow
Freq
Operation
PMAX
Trigger
PAW
105