Supply From A Piped Medical Gas System; Hanging Oxylog 3000 On Standard Rail Systems - Dräger Oxylog 3000 Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000:
Table of Contents

Advertisement

Assembly
5 Rotate the cylinder valve (C) slowly and open
fully.
CAUTION
Do not connect flow control valves or flowmeters
in the gas supply to Oxylog 3000.
The ventilator could malfunction!
WARNING
Always check the O
2
before use, to prevent insufficient oxygen
supply during use.

Hanging Oxylog 3000 on standard rail systems

The Oxylog 3000 can be hung on various rail sys-
tems measuring up to 35 mm diameter by means of
the claw.
– Ensure that the rail is completely inserted in the
claw.
– To ensure optimal functioning of the claw, a dis-
tance of at least 35 mm between rail and wall is
required.
36
pressure of cylinder

Supply from a piped medical gas system

A
1 Connect the O
medical gas hose (A) to the
2
Oxylog 3000.
2 Connect the gas hose (B) to the O
until the supply of O
CAUTION
The Oxylog 3000 is only held by its own weight
when hung on a bar or rail. The Oxylog 3000 must
be secured additionally when being transported,
otherwise vibrations may cause accidental
dislodgement.
Instructions for Use Oxylog 3000 SW 1.n
B
terminal unit
2
is confirmed.