Dräger Oxylog 3000 Instructions For Use Manual page 97

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000:
Table of Contents

Advertisement

A
C
4 Disconnect the flow sensor (A) from the
breathing valve.
WARNING
Do not twist or use force when disconnecting
the flow measuring hoses from the flow
sensor nozzles. This can damage the flow
sensor.
Risk of malfunction.
5 Carefully detach the flow measuring hoses (B)
from the flow sensor.
Pull in the axial direction of the hose nozzles.
6 Detach the angled connector (C) from the flow
sensor.
WARNING
Do not allow any objects to enter the flow
sensor to prevent risk of malfunctions.
Do not purge with compressed air. The wind
vane inside may be damaged and cause
measuring errors!
7 Detach the ventilation hose (D) from the
breathing valve.
Instructions for Use Oxylog 3000 SW 1.n
D
B
B
Cleaning, Disinfection and Sterilization
Breathing valve, disassembly
A
B
8 Turn the cover (A) about 90
counterclockwise to unlock the cover.
9 Remove the silicone diaphragm (B).
WARNING
Do not disassemble breathing valve any
further.
Do not allow any objects to enter the housing
of the breathing valve to prevent risk of
malfunctions!
Do not damage the silicone diaphragm and
other parts.
Do not purge with compressed air. The
silicone diaphragm may be damaged or
misplaced. Malfunction of the breathing valve
can endanger the patient.
Risk of CO
rebreathing.
2
o
97