Maintenance
Maintenance intervals
Inspection
examination of actual condition.
Service
measures to maintain specified
condition.
Repair
measures to restore specified
condition.
Maintenance inspection, service, and repair,
where necessary.
Preventive
maintenance measures at regular
Maintenance
intervals.
CAUTION
In order to avoid malfunctioning of the device,
maintenance must be carried out by properly
trained service personnel.
CAUTION
Have the ventilator inspected and serviced at
regular two-year intervals.
Keep a record on all preventive maintenance.
WARNING
Clean and disinfect the device or device parts
before each maintenance step – and also
when returning for repair to prevent risk of
infection.
Task
1)
Replace dust filter.
Replace internal battery.
Device inspection and maintenance.
1) The dust filter can be treated as household waste.
2) Refer to "Technical Data" section for the battery operating time.
104
Frequency
Every two years.
– Every two years
– When the battery no longer
remains charged for the speci-
fied operating time
Every two years.
Performed by
Trained service personnel.
Trained service personnel.
2)
.
Trained service personnel.
Instructions for Use Oxylog 3000 SW 1.n