Monitoring; Expiratory Minute Volume Mv; Airway Pressure Paw; Apnoea Alarm Time Tapnoea - Dräger Oxylog 3000 Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000:
Table of Contents

Advertisement

Technical Data

Monitoring

Expiratory minute volume MV

Alarm, upper alarm limit
Alarm, lower alarm limit

Airway pressure Paw

Alarm, upper alarm limit
Alarm, lower alarm limit

Apnoea alarm time TApnoea

Alarm
116
Range of settings
Range of settings
Range of settings
Range of settings
When the upper alarm limit has been exceeded.
2 to 41 L/min
When the level drops below the lower alarm limit.
0.5 to 40 L/min
When value "Pmax" is exceeded.
20 to 60 mbar
When the pressure difference between inspiratory
and expiratory phases is less than 5 mbar.
Or
If the set pressure level is not attained.
When respiratory activity is no longer detected.
15 to 60 s, can be set in 1 s increments.
Instructions for Use Oxylog 3000 SW 1.n