Assemble The Reusable Hose System; Breathing Valve Assembly - Dräger Oxylog 3000 Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000:
Table of Contents

Advertisement

Assemble the reusable hose system

– Parts must always be sterilized before use!

Breathing valve assembly

WARNING
The rubber disc (A) in the housing may not be
removed, damaged or bent, otherwise the
valve will not work properly and will endanger
the patient.
Risk of CO
rebreathing.
2
Instructions for Use Oxylog 3000 SW 1.n
A
1 Place the diaphragm (B) in the breathing valve.
2 Fit the cover (A) and turn it approximately 90
3 Push the flow sensor (C) into breathing valve.
4 Push the angled connector (D) onto the flow
WARNING
Always use an angled connector. If the angled
connector is not used, the minute volume may
be measured incorrectly.
A
B
C
D
Ensure that it is inserted correctly.
clockwise to secure into position.
Note the preferred position as indicated by the
groove.
sensor.
Assembly