Gas Supply - Dräger Oxylog 3000 Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000:
Table of Contents

Advertisement

Technical Data
AC/DC power pack
DC/DC converter

Gas supply

1) NIST = Non Interchangeable Screw Thread Connection
2) DISS = Diameter Index Safety Systems
3) N-F = French standard
118
Temperature range
Protection class to
EN 60601
Input
Output
Temperature range
Input
Output
O
service pressure
2
Supply gas
O
inlet connection
2
Gas cylinders and
pressure reducers
Pressure reducer
Gas consumption for
internal control
Accuracy of gas
consumption indication
o
o
–20
C to 50
C
Class II, the earthing is used for EMC purposes.
100 to 240 V~/ 50 to 60 Hz / 0.6-0.3 A
19 V ±0.5 V / 2.1 A
To isolate the ventilator system from mains,
disconnect the power cord from the wall connector.
o
o
–20
C to 50
C
10 / 24 / 28 V DC; 9 A / 4.5 A / 4 A
19 V ±0.5 V / 2.1 A
From a pipeline system or from a medical gas
cylinder.
300 kPa –10 % to 600 kPa at 100 L/min
Medical oxygen or medical air.
either:
1)
NIST
to EN 739, or
2)
DISS
to CGA V5-1989, or
3)
N-F
S90-116
WARNING
Only use medical grade oxygen or medical grade
air that is dry and free from dust and oil.
Contaminated gas can cause lung damage and
device malfunction.
WARNING
Only use compressed gas cylinders and
pressure reducers, which comply with all
applicable regulations and have been approved.
Must have a vent valve on the output side to limit the
delivery pressure to approximately 1000 kPa in the
event of a fault.
Average 0.5 L/min
15 % or ±1 L/min, whichever is greater.
Instructions for Use Oxylog 3000 SW 1.n