Dräger Oxylog 3000 Instructions For Use Manual page 68

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000:
Table of Contents

Advertisement

Alarms
Setting alarm limits automatically
WARNING
After using the function Auto alarm limits:
check if the new alarm limits are appropriate
for the patient.
Risk of hypoventilation.
The function Auto alarm limits sets the alarm
limits on the basis of the following actual measured
values at the time of activation:
MV
: Measured value MV +2 L/min
MV
: Measured value MV –2 L/min
fspn
: Measured value fspn +5/min, with
a minimum of 10/min.
This automatic selection of alarm limits is
performed only once, when confirmed, via the
rotary knob. The alarm limits refer to the current
measured values for MV and fspn.
68
Instructions for Use Oxylog 3000 SW 1.n