Messages in the information window
(Numerical examples)
Message
f = 12 per min or
VT = 800 mL
I:E = 1 : 1.5
Flow = 15 L/min
f = 12 per min or
VT = 800 mL
Tinsp= 0.7 s
Flow = 35 L/min
I:E = 1 : 1.5
Texp= 2 s
Confirm PEEP above 10 mbar ?
Gas consumption = 10 L/min
(Battery capacity)
Pinsp>=PEEP + 3 mbar
PASB = 22 mbar
Instructions for Use Oxylog 3000 SW 1.n
Cause
Change in Tinsp, f or VT in ventila-
tion mode SIMV.
Change in I:E, f or VT in ventilation
mode IPPV (CMV), IPPVAssist
(CMVAssist).
Change in Tinsp or f in ventilation
®
mode BIPAP
(PCV+).
PEEP >10 mbar has been set but
not confirmed.
Standard display in information
window for the current gas
consumption.
Standard display in information
window for the current battery
capacity.
Set PEEP + 3 mbar > Pinsp
Change in ΔASB or PEEP.
Problem Solving
Explanation/Remedy
The required setting of
PEEP >10 mbar is only possible
when confirmed via the rotary
knob.
Set Pinsp > PEEP + 3 mbar
PASB is the absolute pressure
resulting from PEEP + ΔASB.