Customer Service Mode

WARNING
Ventilation is not possible in customer service
mode.
In customer service mode, the ventilator performs
function tests, outputs status information and
permits configuration of parameter settings.
Displays in customer service mode appear in
English and cannot be changed to any other lan-
guage.
001
Set startup settings.
002
Select hose type.
003
Set date and time
(Greenwich Mean Time GMT).
004
Set measured values display window. Configure the layout of measured values in the measured
005
Enter activation code.
006
Test buttons and potentiometer.
007
Test loudspeaker, buzzer, LEDs and
display.
008
Display accu and supply data.
009
Display actual technical errors.
010
Display error and info logbook.
011
Display settings logbook.
012
Display language text.
Instructions for Use Oxylog 3000 SW 1.n
Configure start-up settings, restore manufacturer's
default settings.
Determine which ventilation hose system is used
(reusable or disposable hose system).
Set date and time.
values window or restore manufacturer's default settings.
Enter the activation code for options.
Check for correct functioning of keys and controls.
Check for correct functioning of loudspeaker, buzzer,
LEDs and display.
Display battery data and condition of the supply voltage.
Display any active technical errors.
Calibration logbook and technical errors in chronological
order.
Logbook of operating phases and ventilator settings.
Display screen texts in two selectable languages.