HBM T10FH Mounting Instructions page 33

Torque flange
Hide thumbs Also See for T10FH:
Table of Contents

Advertisement

T10FH
Parameter
Calibration Date
Calibration Initials
Calibration Period
(Days)
Measurement
location ID
1)
Typical values for an HBM T10FH/150 kN@m torque flange
Template: HBM Channel Name
Channel name
When the manufacturer creates the Bridge Sensor template,
the physical measured quantity and the physical unit are defined.
The available unit for the particular measured quantity is specified in the IEEE
Standard. For the measured quantity of torque, the unit is "N@m".
Also in the template, the choice has to be made between "Full precision",
"mV/V" and "uV/V" for the accuracy of the characteristic curve of the trans
ducer mapped in TEDS.
The factory setting is "Full Precision", in order to be able to use full digital res
olution. This choice is also recommended to users who program the TEDS
memory themselves.
A2037−4.0 en
Value
Unit
1)
1−Nov−2005
CAL
HBM
730
days
0
Require
Explanation
d user
rights
Date of the last calibration or creation
of the test certificate (if no calibration
carried out), or of the storage of the
TEDS data (if only nominal (rated)
values from the data sheet were used).
Format: day−month−year.
Abbreviations for the months: Jan, Feb,
Mar, Apr, May, Jun, Jul, Aug, Sep, Oct,
Nov, Dec.
CAL
Initials of the calibrator or calibration
laboratory concerned.
CAL
Time before recalibration, calculated
from the date specified under
Calibration Date.
USR
Identification number for the measuring
point.
Can be assigned according to the
application. Possible values: a number
from 0 to 2047. If that is not enough,
the HBM Channel Comment template
can also be used
for this purpose.
T10FH/150 kNm
33