HBM T10FH Mounting Instructions page 32

Torque flange
Hide thumbs Also See for T10FH:
Table of Contents

Advertisement

32
Area 3:
Data specified by the manufacturer and the user are contained in this area:
For the T10FH torque flange, HBM has already described the Bridge Sensor
and Channel name templates.
Additional templates, such as the Signal Conditioning template, can also
be described by the user.
Template: Bridge Sensor
Parameter
Transducer
Electrical Signal
Type
Minimum Torque
Maximum Torque
Minimum
Electrical
Value
Maximum
Electrical Value
Mapping Method
Bridge type
Impedance of
each bridge
element
Response Time
Excitation Level
(Nominal)
Excitation Level
(Minimum)
Excitation Level
(Maximum)
1)
Typical values for an HBM T10FH/150 kN@m torque flange
HBM
Value
Unit
1)
Bridge
Sensor
0.000
N@m
150000
N@m
0.0000m
V/V
1.8245m
V/V
Linear
Full
1550+−100
Ohm
1.0000000u
s
5.0
V
2.5
V
12.0
V
Require
Explanation
d user
rights
ID
CAL
The physical measured quantity and
unit are defined when the template is
created, after which they cannot be
changed.
CAL
CAL
The difference between these values is
the sensitivity according to the HBM
test certificate or from the calibration.
CAL
This entry cannot be changed
ID
The bridge type. "Full" for a full bridge.
ID
Input resistance according to the HBM
data sheet.
ID
Of no significance for HBM transducers
ID
Nominal (rated) excitation voltage
according to the HBM data sheet
ID
Lower limit for the operating range of
the excitation voltage according to the
HBM data sheet
ID
Upper limit for the operating range of
the excitation voltage according to the
HBM data sheet
T10FH
A2037−4.0 en

Advertisement

Table of Contents
loading

Table of Contents