Vibration
Flammability
Characteristics of the options
Power meter
option
VFL
Bluetooth
Option
The 2000 Platform resists the following vibration tests:
–
Complete test comprising 6 cycles along each of the x, y and z axes.
–
One cycle of 5 to 200 Hz and back to 5 Hz with a sweep duration of
one minute/octave.
–
3 mm amplitude displacement test, for the range 5 Hz to 15 Hz.
–
3g acceleration test for the range 16 Hz to 200 Hz.
The 2000 Platform housing (in ABS, type V0) does not propagate fire.
Specifications given for 25°C, after 20 minutes stabilization time and after
zero setting.
–
Wavelength range: 800 to 1650 nm in steps of 1 nm
–
Calibrated wavelengths: 850 / 1310 / 1490 / 1550 / 1625 nm
–
Accuracy at calibrated wavelengths: ± 0.2 dB (at -30 dBm)
–
Input power range : -60 dBm to +10 dBm
–
Maximum resolution: 0.01 dB / 0.01nW
–
Measurement range:+5 to -50 dBm (-45 dBm from 800 to 1250 nm)
–
Linearity within the measurement range: ± 0.2 dB
–
Wavelength: 650 nm
–
Length of fiber: up to 5 km
–
Class 2 laser (standards EN60825-1 and FDA21 CFR Part
1040.10).
–
Bluetooth Option
–
Class 2
–
Range: up to 10 meters
User Manual
790000002/00
Chapter 9 Technical specifications
Characteristics of the options
73