If all the targets are not correctly touched, the touchscreen cannot be
used. Otherwise, the touchscreen may be used directly.
Target
Fig. 53
Touchscreen calibration
If you meet problems during the Platform functioning, or if the battery
does not charge anymore when plugged, this may require the battery to
be replaced.
CAUTION
Battery is not interchangeable in the field. It must be replaced
exclusively for maintenance purpose.
To access the battery of the 2000 Platform, proceed as follows:
1
Switch off the instrument and disconnect the mains supply.
2
Turn the instrument face down on the work surface.
3
If any, remove the module installed (see
page
14).
4
Remove the battery door
5
Pull the battery connector from its housing, to disconnect it from the
base, taking care not to damage the connector into which it is
plugged.