GRAPHTEC GL500 User Manual page 98

Midi logger dual
Table of Contents

Advertisement

• Stop
Function: Specify the trigger condition that must be met in order to stop measurement.
Off: No trigger is used. Instead, measurement is stopped by pressing the START/
STOP key.
Amp: Measurement is stopped when the level trigger condition is satisfied.
Note: See the CHECKPOINT on page 3-19 for more details on the Level Trigger.
Date: Measurement stops at the specified time.
Note: If Trigger Repeat has been specified, perform the date and time settings.
Ext.: Measurement is stopped when a signal is input from an external trigger
terminal.
Time: Set the amount of time from the start until the end of measurement. The time
can be specified in 1-s increments.
Event: Measurement stops when all the memory blocks are filled with event capture
data.
Trigger Repeat: Insert a check to enable repeated measurement.
(4) Amp Trigger Settings
Select CH: Select the channel display group.
1 to 8, 9-16
• Start Trigger
CH: Select the channels for which you want to specify a start trigger.
Function: Specify the Trigger Start condition.
Off: Click the "Start" key to start measurement.
Hi: The trigger is activated when the signal input rises to (or exceeds) the specified
upper limit.
Lo: The trigger is activated when the signal input falls to (or goes below) the
specified lower limit.
Win In: The trigger is activated when the signal input goes within (or is within) both
Win Out: The trigger is activated when the signal input goes outside (or is outside)
Upper: Input the signal input value for the upper limit.
Lower: Input the signal input value for the lower limit.
Combination: Select either AND (all conditions met) or OR (any one of the conditions is met)
for the channels for which the Start Trigger has been specified.
• Stop Trigger
CH: Select the channels for which you want to specify a stop trigger.
Function: Specify the Trigger Stop condition.
Off: Click the "Stop" key to stop measurement.
Hi: The trigger is activated when the signal input rises to (or exceeds) the specified
upper limit.
Lo: The trigger is activated when the signal input falls to (or goes below) the
specified lower limit.
Win In: The trigger is activated when the signal input goes within (or is within) both
Win Out: The trigger is activated when the signal input goes outside (or is outside)
Combination: Select either AND (all conditions met) or OR (any one of the conditions is met)
for the channels for which the Stop Trigger has been specified.