GRAPHTEC GL500 User Manual page 56

Midi logger dual
Table of Contents

Advertisement

Settings and Measurement
File Name ........................... This parameter is displayed when PC card is selected as the capture
<Trigger Settings>
Start
Source ................................ Specifies the trigger condition that must be met in order to initiate
3-18
destination.
File Type: This sets the file format for the data to be saved.
• GBD: Binary format
• CSV: Excel format
Name Type: Either Auto or User can be selected.
Folder: The currently selected folder is displayed. Settings cannot be
made here.
File name (User): A new folder and a new file are created here.
measurement.
Off:
No trigger is used. Instead, measurement is initiated by
pressing the START/STOP key.
Amp→Amp Settings (Level Trigger)
The start of measurement is triggered when the level trigger
condition is satisfied.
Note: Set the Level condition as the level value of the range (voltage,
temperature) set for the specified channel.
Date: Measurement starts at the specified time.
Note: Since measurement starts at the specified time, make sure that the
GL500 time settings are correct. Moreover, if Repeat has been set to
On, data capture will start at the specified time each day.
Ext.: The start of measurement is triggered when a signal is input
from an external trigger terminal.