Pulse ................................... The signals that can be input to the pulse input terminals for data
(b) Line Color
Use the Color Settings window to change the display colors for
each channel.
Other Misc. settings
For voltage settings, zero point adjustment can be performed
automatically, and the zero point reset.
(c) Perform Auto Zero ADJ
Sets the current input voltage as the zero point voltage value.
(d) Reset Auto Zero ADJ
Cancels the zero point voltage value, and displays the input
voltage.
(e) Set Zero Point as
Displays the current zero point voltage value.
processing are Revol., Counts, and Inst.
• Mode: Select from Revol., Counts, and Inst.
Revol: Counts the number of pulses per second, and
displays the values multiplied by 60 as rpm values.
Counts: Displays a count of the number of pulses for each
sampling interval from the start of measurement.
Inst.:
Displays the number of pulses for each sampling
interval.
• Color: Use to change the display colors for each channel.
• Range: Can be set for each of the modes.
Mode
Revol.
5, 50, 500, 5k, 50k RPM/F.S
Counts
5, 50, 500, 5k, 50k C/F.S.
Inst.
5, 50, 500, 5k, 50k C/F.S.
• Filter: Set the filter to On when there is likely to be noise in the input.
Off: Disables the filter
On: Enables the filter
Settings and Measurement
Range
Maximum number
of pulse inputs
50k/s
50k/sampling interval
50k/sampling interval
3-15